矢量网络分析仪校准误差分析
Influence of Nonideal Calibration Items on VNA Measuring
-
摘要: 本文针对SOLR 双端口校准方法,分析了校准件不完善——开路器、短路器校准位置不明确及校准件负载不理想对双端口校准结果的影响,并推导出校准件实际值和理想值的偏差与校准结果之间的数学关系,结合仿真数据对推导结果进行了验证。Abstract: The influence of nonideal calibration Items on short-open-load-reciprocal (SOLR)calibration method to calibrate a four sampler VNA is considered in this paper .The expressions for the deviations of the error terms and of the measured S-parameters of two-port test objects from their true values are presented.Moreover,the data from simulation validate the expressions at the end of this paper.
下载: